ANSWERS OF CLARIFICATION REQUESTS OF TENDER
EUROPEAID/120263/D/S/TR/BIS
CLARIFICATION NO 1
"Support to the Turkish Conformity Assessment
Bodies and Ministry of Industry and Trade
in the Implementation of some New Approach-Supply of Equipment for
CABs"
PLEASE NOTE THAT CORRIGENDUM RELATING TO THIS CLARIFICATION WILL BE
PUBLISHED VERY SOON.
The following
alterations and/or corrections are made to the Tender Dossier
GENERAL QUESTIONS
ANSWER NO 1
The
effective and continuous link with the economy of the state concerned means
that a legal entity, company or partnership having only its statuary office in
one of the eligible states must be legally well-established in that state with
demonstration of economic activities over a certain period of time.
LOT 2- ELECTROMAGNETIC COMPATIBILITY
(EMC) - TESTING EQUIPMENT
QUESTION NO 1
RE 2.1.1. The chamber dimensions will
allow fully compliant immunity testing and pre-compliant emission testing, is
this acceptable? If the chamber were to be 1.5m longer then fully compliant
emissions would also be possible. From Corrigendum No3 – “Site validation: The
normalized site attenuation curve should not deviate from the theoretical ANSI
C 63.4 1992, CISPR 16, and EN 50147-2 site attenuation values by more than
±3.8dB.
-The Space
NSA correlation should not deviate from the theoretical site attenuation values
(similar procedure as described ANSI C 63.4 1992) by more than ±3,5dB.
-The supplier
must guarantee –0dB to +6dB field strength uniformity over a 1.5m x 1.5m
vertical area located 0.8m above the floor as described in EN 61000-4-3+A1”.
Given the
size of the chamber, it is not possible to meet this over the full frequency
range, a relaxation to +/-5dB should be given below 100MHz.
ANSWER
NO 1
The
size of the chamber is not changed.
QUESTION NO 2
RE 2.1.4. There is no power source specified
for this test. In order to make measurements to EN61000-3-2/3 a clean power
supply is necessary as the public mains is unlikely to be suitable. Should the
power source be added to the tender response?
ANSWER NO 2
No, power source is not needed.
QUESTION NO 3
RE2.1.5. One injection clamp can cover
the whole range, should calibration adapters be included?
ANSWER NO 3
Yes, calibration adapters should be added.
QUESTION NO 4
RE 2.1.10. Should we also offer a
calibration fixture, termination and monitor probe to comply with EN61000-4-6?
ANSWER NO 4
No, Calibration fixture,
termination and monitor probe to comply with
EN61000-4-6 should not
be offered.
QUESTION NO 5
It is unclear as to whether a full system
is required or just the equipment list. The list specified is insufficient to
do an EN61000-4-3 or any other radiated immunity test, the following are also
required and should they be added to the tender response?
a. Immunity antennae, it is not recommended to connect a 300W
amplifier at 3GHz to a 300W rated
emission antenna. They should cover 80MHz to 3GHz, log periodic and horn.
b.
Directional couplers to sample the amplifier power, matched to the specified amplifiers.
c. Power
meter to measure the amplifier power. 10k-3GHz, +10 to -50dBm.
d. Field
measurement probe to measure and calibrate the generated field, 10k-4GHz.
ANSWER
NO 5
No,
the above listed equipment should not be added to the tender response.
QUESTION NO 6
It is also unclear as to whether the
system should be automatic or manual; if automatic then the following should be
added to the tender response?
a. Control
software for emissions and immunity
b. Computer
and interfaces
c. Switch
system
d. Cables and
connectors
e. Rack
housing
ANSWER
NO 6
The above mentioned equipment was already
listed in the tender dossier. The offer
should include the necessary equipment for automatic system.
LOT 3- MACHINERY – TESTING &
SUPPORT EQUIPMENT
QUESTION NO 1
180 days delivery is almost
impossible to combine for
ANSWER NO 1
The project period has
been extended to 270 calendar days for this lot only. Please see corrigendum no
1
QUESTION NO 1
4.1.1.1.
ICP-MS-HPLC System :
Detection System: Please
advise if all pulse counting (digital)
longer lifetime detector providing 9 orders of magnitude linear dynamic range (comparing with older design detectors covering 8 orders of magnitude linear dynamic range) and having simpler operation ( no need
for cross calibration as it operates in pulse counting mode)
than analog/pulse detectors (which require cross calibration
function to combine the calibration graphs of analog and pulse counting modes
having different linearities for the same element) is also acceptable?
ANSWER NO 1
QUESTION NO 2
4.1.6.
Microwave Muffle Furnace:
Capacity & Operating Temperature: There are two furnace sizes available for our ashing
systems: The High Temperature Chamber
ramps up to 1200 ºC and will hold up to 8 (25 ml) crucibles with total
chamber capacity of 1.8 liters.
and
The High
Capacity Chamber ramps up to 1000
ºC and holds up to 15 (25 ml)
crucibles with total chamber capacity of 5.0 liters.
Please clarify which chamber to be quoted . Or
alternatively both chambers can be used in the same furnace so both chambers
can be quoted. Please advise.
ANSWER NO 2
Operation temperature of up
to 1200 ºC is required; High Temperature Chamber that ramps up to
1200 ºC with total chamber capacity of 1.8. liters is acceptable.